Changes between Version 9 and Version 10 of WikiStart
- Timestamp:
- Mar 17, 2017, 12:24:02 PM (8 years ago)
Legend:
- Unmodified
- Added
- Removed
- Modified
-
WikiStart
v9 v10 42 42 and 11mA on the 2.45A scale. 43 43 We also find that, assuming an ideal power supply without noise or ripple, 44 we have a +/-7 unit error on the ADC's output . This is less than 0.18% of the45 full range (or +/- 4.4mA on the 2.45A range input). Below are plots for measures on the 0.61A scale.44 we have a +/-7 unit error on the ADC's output for current values. 45 This is less than 0.18% of the full range (or +/- 4.4mA on the 2.45A range input). Below are plots for measures on the 0.61A scale. 46 46 47 47 [[Image(0mA.gif, 50%, align=left, title="ADC values for 0mA",longdesc="ADC values for 0mA")]] … … 53 53 at most 4, which is within the noise. 54 54 [[Image(i_o.gif, 50%, align=right, title="current to ADC outut",longdesc="current to ADC output")]] 55 56 We repeated the process for the tension inputs. Here we have a +/-5 unit error on the ADC's output for voltage inputs. This is less than 0.13% of the full range (or +/- 29mV on the 23.88V range). Below are plots for measures on the 6V scale. 57 [[Image(0.171V.gif, 50%, align=left, title="ADC values for 0.171V",longdesc="ADC values for 0.171V")]] 58 [[Image(5.996V.gif, 50%, align=right, title="ADC values for 5.996V",longdesc="ADC values for 5.996V")]]